Carbon based resistive strain gauge sensor fabricated on titanium using micro-dispensing direct write technology
Date
Authors
Advisors
Journal Title
Journal ISSN
ISSN
Volume Title
Publisher
Type
Peer reviewed
Abstract
Carbon based resistive strain gauge sensor suitable for medical implant technology has been directly fabricated on a titanium test-plate using Micro-Dispensing Direct Write (MDDW) technology. A 3.5µm biocompatible dielectric layer of parylene C was initially coated on the titanium test-plate. Commercially available screen-print carbon conductive paste was deposited on the parylene C and cured at 80oC for 3 hours; this was to ensure the physical properties and chemical integrity of the parylene C layer was maintained, whilst meeting the electrical conductivity curing requirements for the carbon tracks. The novel integrated strain sensor was experimentally tested and found to have a gauge factor of 10 making it approximately 5 times more sensitive than a commercially available metal foil strain gauge glued to the same titanium plate.